Paper

Home >> Paper

Improving Information Collection Environment of Magnetocardiography Device Through Adaptive Interference Suppression Strategy

Release time:2026-09-13| Hits:

Impact Factor:7.9

DOI number:10.1109/TASE.2025.3567988

Affiliation of Author(s):仪器科学与光电工程学院

Teaching and Research Group:大科学装置研究院

Journal:IEEE Transactions on Automation Science and Engineering

Abstract:Multiple frequencies high-amplitude magnetic interference in the environment can adversely affect the detection of cardiac magnetic field signals inside magnetocardiography device. Adaptive multi-resonance control (AMR) is suitable for suppressing such interference, but the stability of active magnetic field compensation system (AMC) will be destroyed by the influence of the time delay of the sensor. To address this problem, this paper proposes a interference suppression method that combines an improved adaptive multi-resonance controller (IAMR) with proportional differential (PD) and linear extend

Co-author:Haifeng Zhang, Zhihui Hong, Shiqiang Zheng

First Author:Pengtao Tian

Indexed by:Journal paper

Discipline:工学

First-Level Discipline:Instrumentation Science and Technology

Document Type:J

Volume:22

Page Number:15264-15273

ISSN No.:1545-5955, 1558-3783

Translation or Not:no

Date of Publication:2025-01-01