Impact Factor:7.4
DOI number:10.1109/TIE.2025.3613643
Affiliation of Author(s):仪器科学与光电工程学院
Teaching and Research Group:大科学装置研究院
Journal:IEEE Transactions on Industrial Electronics
Key Words:Magnetic field measurement; Magnetic shielding; Magnetic fields; Magnetic resonance imaging; Coils; Magnetic noise; Harmonic analysis; Magnetic resonance; Active magnetic disturbance suppression; composite controller; harmonic disturbances suppression; linear extended state observer (LESO); nonlinear gradient coil; Signal to noise ratio; State estimation
Abstract:High-power devices used in hospitals generate harmonic magnetic field disturbances inside magnetic shielding cylinders (MSCs). These disturbances exhibit axial-gradient spatial distributions, which substantially degrade the imaging quality of magnetocardiography (MCG). To address this issue, a high-precision nonlinear spatial gradient disturbance suppression method is proposed for single-open MSCs. First, a nonlinear gradient coil design is developed to generate variable magnetic field gradients, precisely matching the nonlinear gradient distribution induced by external magnetic disturbances i
Co-author:Zhongxiang Jing, Fengwen Zhao, Xiangyang Ye, Suhang Li, Yunyang Lian, Gang Liu
First Author:Haifeng Zhang
Indexed by:Journal paper
Discipline:工学
First-Level Discipline:Instrumentation Science and Technology
Document Type:J
Volume:73
Issue:2
Page Number:3316-3327
ISSN No.:0278-0046, 1557-9948
Translation or Not:no
Date of Publication:2026-02-01